Non-contact Quantitative Diffuse Reflectance Spectroscopy

Veronica Sorgato, Michel Berger, Charlotte Emain, Anne Koenig, Blandine Roig, Christine Vever-Bizet, Jean-Marc Dinten, Genevieve Bourg-Heckly, Anne Planat-Chretien

Résumé

We validate a non-contact Diffuse Reflectance Spectroscopy (DRS) system as a first stage to approach quantitative multi-spectral imaging technique. The non-contact DRS system with separated illumination and detection paths was developed with different progressive set-ups which were all compared to a well-founded contact DRS system. While quantitation of the absorption coefficient is well achieved with the existing method, the calculation of the scattering coefficient is deteriorated by the non-contact architecture measurements. We have therefore developed an adaptive reference-based algorithm to compensate for this effect.

Citer cet article

Sorgato, V., Berger, M., Emain, C., Koenig, A., Roig, B., Vever-Bizet, C., Dinten, J.-M., Bourg-Heckly, G., & Planat-Chretien, A. (2015). Non-contact Quantitative Diffuse Reflectance Spectroscopy. DIFFUSE OPTICAL IMAGING V, 9538(95380U). https://doi.org/10.1117/12.2183764