proceeding · 01/01/2011
Toward nonlinear endomicroscopy for exploration of the pulmonary airways: preliminary spectroscopic study of human lung tissue using a single-mode fibre
Résumé
We are developing a non-linear fibered endomicroscope for imaging the extracellular matrix collagen and elastin fibrillar networks during bronchoscopy. As a proof of concept, laser pulses at the output of a standard 2 meter long single-mode fibre have been obtained with pulse duration of about 50 fs and pulse energy up to 50 uJ, using a specially designed grism line for the dispersion compensation. With these pulses, we performed a spectroscopic characterization of the non-linear endogenous signal, consisting of two-photon fluorescence and second harmonic generation, and originated from human pulmonary tissue of various thickness, both in forward and backward geometry of signal collection, with excitation at 830 nm.
Citer cet article
Peyrot, D.-A., Lefort, C., Haidar, D.-A., Mansuryan, T., Kruglik, S., Sandeau, N., Louradour, F., & Bourg-Heckly, G. (2011). Toward nonlinear endomicroscopy for exploration of the pulmonary airways: preliminary spectroscopic study of human lung tissue using a single-mode fibre. ENDOSCOPIC MICROSCOPY VI, 7893(789309). https://doi.org/10.1117/12.875440
@inproceedings{Peyrot2011_35,
author = {Peyrot, Donald A. and Lefort, Claire and Haidar, Darine Abi and Mansuryan, Tigran and Kruglik, Sergei and Sandeau, Nicolas and Louradour, Frederic and Bourg-Heckly, Genevieve},
year = {2011},
month = {1},
title = {Toward nonlinear endomicroscopy for exploration of the pulmonary airways: preliminary spectroscopic study of human lung tissue using a single-mode fibre},
booktitle = {ENDOSCOPIC MICROSCOPY VI},
publisher = {SPIE-INT SOC OPTICAL ENGINEERING},
volume = {7893},
number = {789309},
series = {Proceedings of SPIE},
address = {1000 20TH ST, PO BOX 10, BELLINGHAM, WA 98227-0010 USA},
abstract = {We are developing a non-linear fibered endomicroscope for imaging the extracellular matrix collagen and elastin fibrillar networks during bronchoscopy. As a proof of concept, laser pulses at the output of a standard 2 meter long single-mode fibre have been obtained with pulse duration of about 50 fs and pulse energy up to 50 uJ, using a specially designed grism line for the dispersion compensation. With these pulses, we performed a spectroscopic characterization of the non-linear endogenous signal, consisting of two-photon fluorescence and second harmonic generation, and originated from human pulmonary tissue of various thickness, both in forward and backward geometry of signal collection, with excitation at 830 nm.},
note = {Conference on Endoscopic Microscopy VI, San Francisco, CA, JAN 23-24, 2011},
url = {http://www.dx.doi.org/10.1117/12.875440},
doi = {10.1117/12.875440},
issn = {0277-786X},
}
TY - CONF
AU - Peyrot, Donald A.
AU - Lefort, Claire
AU - Haidar, Darine Abi
AU - Mansuryan, Tigran
AU - Kruglik, Sergei
AU - Sandeau, Nicolas
AU - Louradour, Frederic
AU - Bourg-Heckly, Genevieve
PY - 2011
DA - 2011/01/01
TI - Toward nonlinear endomicroscopy for exploration of the pulmonary airways: preliminary spectroscopic study of human lung tissue using a single-mode fibre
BT - ENDOSCOPIC MICROSCOPY VI
VL - 7893
IS - 789309
PB - SPIE-INT SOC OPTICAL ENGINEERING
SN - 0277-786X
AB - We are developing a non-linear fibered endomicroscope for imaging the extracellular matrix collagen and elastin fibrillar networks during bronchoscopy. As a proof of concept, laser pulses at the output of a standard 2 meter long single-mode fibre have been obtained with pulse duration of about 50 fs and pulse energy up to 50 uJ, using a specially designed grism line for the dispersion compensation. With these pulses, we performed a spectroscopic characterization of the non-linear endogenous signal, consisting of two-photon fluorescence and second harmonic generation, and originated from human pulmonary tissue of various thickness, both in forward and backward geometry of signal collection, with excitation at 830 nm.
DO - 10.1117/12.875440
UR - http://www.dx.doi.org/10.1117/12.875440
ER -